Used for first aid and ICU, can be used for pediatric and adult patients.
Physical Specifications
Dimensions
Width:50cm Depth: 42cm Height: 106cm Weight:25Kg
packing Dimensions
Width:59cm Depth: 45cm Height: 112cm Weight:40Kg
Environmental Specifications
Operating Requirements
Temperature: 5 to 40℃ ;
Relative Humidity: less than 80%;
The atmosphere pressure: 70 to 106 kPa ;
Storage/Shipping Requirements
Temperature: -20 to 70℃
Relative Humidity: less than 93%
The atmosphere pressure: 70 to 106 kPa
Electrical Specifications
Input Voltage Requirements
AC Voltage: 220V+/-10%
Frequency: 50HZ+/-2%
Power: 170 VA;
Fuse Specification Information
Quantity: 2
Current Rating: 0.8 A
Voltage Rating: 250 Volts
Type: 5*20 mm RT
Output Voltage Specification Information
AC Voltage: 220V+/-10%,
Frequency: 50HZ+/-2%
Power: 130VA
Pneumatic Specifications
Oxygen Inlet
Pressurized Oxygen: 0.45 MPa ±10% medical oxygen.
IEC Classification
Ordinary equipment, not protected against the ingress of liquids
Not protected/Not suitable for use in the presence of flammable
anesthetic gases.
ClassⅠ, Type B
Performance Specification And Tolerances
Ventilator Parameters | Ranges | Remark |
Controls | ||
Power | ON/OFF | |
Modes: IPPV(CMV) IPPV+SIGH SIPPV(A/C) SIMV MANUAL BREATH | ||
Tidal Volume | 0~1200ml | |
Breath Rate | 5~60 bpm | |
PEEP | 0~1KPa | |
%O2 | 45~90% | |
Sigh | ||
Inspiratory Time | 0.5~3sec | |
Trigger Level | -0.5~1kpa | |
Over Pressure Relief | less than 6kpa |
Performance Specification And Tolerances
Ventilator Parameters | Ranges | Tolerances |
Alarms and Indicators | ||
High Pressure Alarm Limit | 2~6kpa flash, red indicator | +/-0.1kpa |
Low Pressure Alarm Limit | 0. 2~2kpa Red indicator | +/-0.1kpa |
High Minute Volume Alarm Limit | 10~30L flash, Red indicator | +/-5% |
Low Minute Volume Alarm Limit | 1~10L flash, Red indicator | +/-5% |
High Rate Alarm Limit | 30~59bpm flash, Red indicator | +/-5% |
Apnea interval | 15, 30, 45, 60second flash, Red indicator | +/-1second |
Alarm Silence | 60 sec. max | |
Power | Red indicator | |
Insp. phase | Green indicator | |
Exh. phase | Green indicator | |
T.VOL Indicator | Green indicator | |
MIN. VOL Indicator | Green indicator | |
PRESSURE Indicator | Green indicator | |
RATE Indicator | Green indicator |
Performance Specification And Tolerances
Ventilator Parameters | Ranges | Remark |
Monitors | ||
Tidal Volume | 0~2560ml | |
MIN. Volume | 0~99L | |
Total Breath Rate | 0~99 bpm | |
Main Circuit Pressure | -2~10KPa |
What Is Provided
Parts | Number | Specification |
Ventilator | 1 | |
Mobile Table | 1 | |
Humidifier | 1 | |
Breath Circuit | 1 | |
Manipulators | 1 | |
Oxygen Pressure Reducer | 1 | |
Mask | 1 | |
Hand Book | 1 | |
VCD For Operation | 1 |